ООО "ЭЛАСО" / тел.: +38-066-709-75-42 / +38-067-524-82-50 / +38-044-259-49-07

Fluke 59 MAX+




Краткие характеристики

Specifications
Temperature Range -30 °C to 500 °C (-22 °F to 932 °F)
Accuracy (Calibration geometry with ambient temperature 23 °C ±2 °C)
≥ 0 °C:   

±1.5 °C or ±1.5 % of reading, whichever is greater

≥-10 °C to <0 °C:

±2.0 °C

< -10 °C: ±3.0 °C
   
≥ 32 °F:    ±3.0 °F or ±1.5 % of reading, whichever is greater
≥ 14 °F to <32 °F:  ±4.0 °F
< 14 °F: ±6.0 °F
Response Time (95 %) <500 ms (95 % of reading)
Spectral Response 8 mm to 14 mm
Emissivity 0.10 to 1.00
Distance to Spot Ratio

10:1 (calculated at 90 % energy)

Display Resolution 0.1 °C (0.2 °F)
Repeatability (% of reading) ±0.8 % of reading or ±1.0 °C 
(±2.0 °F), whichever is greater  
Power 1 AA IEC LR06 Battery
Battery Life 12 hours with laser and backlight on
Weight 220 g (7.76 oz)
Size (156 x 80 x 50) mm (6.14 x 3.15 x 2) inches
Operating Temperature 0 °C to 50 °C (32 °F to 122 °F)
Storage Temperature -20 °C to +60 °C (-4 °F to 140 °F), (without battery)
Operating Humidity 10 % to 90 % RH non-condensing @ 30 °C (86 °F)
Operating Altitude 2000 meters above mean sea level
Storage Altitude 12,000 meters above mean sea level
Drop Test 1 m
Ingress Protection Rating IP40 per IEC 60529
Vibration and Shock IEC 68-2-6 2.5 g, 10 to 200 Hz, IEC 68-2-27, 50 g, 11 ms
Compliance EN/IEC 61010-1
Laser Safety FDA and EN 60825-1 Class II
Electromagnetic Compatibility 61326-1 EN 61326-2